The image guided interventional procedure (IGIP) survey aims to determine diagnostic reference levels (DRLs) for a number of common interventional and diagnostic angiography or fluoroscopic procedures:

  • Diagnostic coronary angiogram (no significant disease)
  • Diagnostic cerebral angiogram 
  • Barium swallow
  • Gastrografin swallow
  • Percutaneous coronary intervention (PCI, single lesion)
  • Endovascular aortic repair (EVAR)
  • Pelvic embolisation
  • Tunnelled line insertion.

The procedure definitions are explained on the survey protocol data sheets. Participants choose which protocols they wish to submit and for statistical accuracy data for 30 patients (50 for cerebral angiogram) per procedure are requested.

To undertake the survey, participants must first register with ARPANSA. Registered participants will then be emailed Excel survey spreadsheets for completion and email return. Additional copies of the worksheets, for surveying more than one room, may be generated by saving the workbook as a new document.

Each compliant submission will result in a dose report being sent to the participating facility. All submitted data and reports will be strictly confidential between ARPANSA and the facility.

No identifying patient information is required or sought. This means that human research ethics approval is NOT required.

Survey metrics definitions

SexM or F
AgeTo nearest number of years in years
WeightTo nearest kilogram in kg
Fluoroscopy timeTotal screening beam on time (does not include run time) in the time format selected in the green ‘Unit Settings’ area
Fluoro frame rateScreening frame rate in frames per time unit as selected in the green ‘Unit Settings’ area
DA timeTotal acquisition run time in seconds
DA frame rateAcquisition frame rate during angiographic runs in frames per time unit as selected in the green ‘Unit Settings’ area
DA FramesTotal number of angiographic frames acquired, does not include digital screening frames
DAPTotal DAP (KAP) value acquired for both screening and angiographic runs in the units selected in the green ‘Unit Settings’ area
Reference doseSystem provided reference dose (for an isocentric interventional fluoroscope, the reference point is located 15 cm from the isocenter toward the X-ray tube), in the units selected in the green ‘Unit Settings’ area
Access site

(Coronary angiography or PCI single lesion)
Catheter insertion via Radial, Femoral or Other approach (R, F or O)

(Tunnelled line)
Catheter insertion area (Left jugular vein, Right jugular vein or Other)

Left Ventriculogram

(Coronary angiography)
Was a left ventriculogram performed for the patient? Tick box for 'Yes'

Entire Oesophagus

(Gastrografin swallow)
Is the entire oesophagus being imaged? (Yes or No)

No. of vessels

(Cerebral angiography)
Number of vessels examined (1 - 6)

(Pelvic embolisation)
Number of vessels embolised

No. of RA acquisitions

(Cerebral angiography)
Number of rotational acquisitions (spins) performed (0 - 6)

No. of catheter changes

(Pelvic embolisation)
Number of catheter changes (complexity metric)

Procedure sub-type

(Pelvic embolisation)
Further classification of procedure (Uterine, Gonadal, Prostate or Other)